LIFEPAK CR2 Pads & Battery: What You Need and When to Replace

The LIFEPAK CR2 is one of the simplest AEDs to maintain, for one reason: it uses a single electrode set for both adults and children, and its pads and battery run on the same four-year clock. No pediatric pads to track separately, no staggered replacement schedule. Here is what your unit takes.

Part numbers at a glance

One electrode set covers every patient

Most AEDs require you to buy, store and track a separate set of pediatric pads. The CR2 does not. The QUIK-STEP adult/child electrodes 11101-000021 are rated for all patients, and pediatric rescues are handled by pressing the Child Mode button on the device rather than by swapping hardware.

Two practical benefits:

  • Nothing to forget. The most common finding in an AED inspection is an expired set of pediatric pads that nobody has looked at in three years. On the CR2 that failure mode does not exist.
  • Nothing to fumble. In a real paediatric emergency, a rescuer presses a button instead of hunting for a second cartridge under pressure.

The pad kit ships with one electrode set, an electrode pads cover and replacement instructions.

The four-year cycle

This is the CR2's other maintenance advantage. Pads are 4-year electrodes and the battery carries a 4-year standby life when installed in a device that has not been used. A new unit ships with both, which means they expire at roughly the same time — so you place one reorder every four years rather than juggling two schedules.

That alignment is why the CR2 suits a bundled reorder better than almost any other AED. Our CR2 Maintenance Package comes in two tiers:

  • Basic — one CR2 battery + one pair of universal electrode pads
  • Standard — one CR2 battery + two pairs of universal electrode pads

Whichever you choose, we contact you 90, 60 and 30 days before they are due.

Battery specifications

Part number 11141-000165
Chemistry Lithium manganese dioxide (Li/MnO₂), non-rechargeable
Rating 12.0 V · 4.7 Ah · 55.8 Wh
Capacity approx. 800 minutes of "on" time, or 166 discharges at 200 J
Standby life 4 years installed in an unused device
Warranty 1 year installed

Where you store it matters

The CR2 battery has published environmental limits that are easy to breach without realising:

  • Operating: 0–50°C (32–122°F)
  • Long-term storage: 0–25°C (32–77°F)
  • Short-term storage: −30–60°C (−22–140°F), for up to one week
  • Humidity: 5–95% non-condensing

Note how narrow the long-term window is. A CR2 living in a vehicle boot, an unconditioned warehouse or an outdoor cabinet through a hot summer spends months above 25°C — within the short-term range, but well outside what the battery is rated to sit in permanently. If that describes your installation, expect real-world standby life shorter than four years, inspect more often, and consider a temperature-rated cabinet.

Which CR2 do you have?

The CR2 ships in several connectivity configurations, and it changes which features you have — though not which consumables you buy. Pads and battery are identical across all of them.

Per the manufacturer's specification, cprINSIGHT analysis (which lets the device analyse the heart rhythm while compressions continue), bilingual switching and remote readiness monitoring are features of the connected Wi-Fi models. The USB-only version does not include them. If you are unsure which you have, the label on the back of the device identifies the configuration.

What to replace after a rescue

Replace the electrode pads after any deployment, even if no shock was advised. The battery is more nuanced on the CR2 than on most units: it is rated for roughly 166 discharges at 200 J, so a single rescue does not exhaust it — but the CR2 escalates from 200 J to 300 J and then to 360 J on subsequent shocks, so a long resuscitation draws considerably more than the headline figure suggests. After any multi-shock event, check the readiness indicator and replace the battery if it is anywhere near the end of its four-year window.

Frequently asked questions

Does the LIFEPAK CR2 need separate pediatric pads?

No. The CR2 uses QUIK-STEP adult/child electrodes, part 11101-000021, which are rated for all patients. Paediatric rescues are handled by pressing the Child Mode button on the device rather than changing pads.

How often do LIFEPAK CR2 pads and battery need replacing?

Both run on a four-year cycle. The QUIK-STEP electrodes are 4-year pads and the battery has a 4-year standby life, so they generally come due at the same time and can be reordered together.

What battery does the LIFEPAK CR2 use?

Part number 11141-000165, a non-rechargeable lithium manganese dioxide battery rated 12.0 V, 4.7 Ah, 55.8 Wh, providing roughly 800 minutes of operating time or 166 discharges at 200 joules.

Can I use LIFEPAK CR Plus pads in a CR2?

No. The CR2 uses QUIK-STEP electrodes, part 11101-000021, while the CR Plus and LIFEPAK Express use their own Charge-Pak and electrode system. The two are not interchangeable.

Reorder for your LIFEPAK CR2

Genuine Physio-Control pads and batteries — and because the CR2 runs both on the same four-year clock, one bundled order covers you until the next cycle.

CR2 Maintenance Package → | Shop CR2 Batteries → | Call 888-541-2337

Related

Last reviewed: 18 August 2026. Part numbers and specifications checked against our live catalogue.

AED Professionals: A General Medical Devices, Inc. company

348 W. Colfax Street, Palatine, IL 60067

info@aedprofessionals.com 847-202-3233

Subscribe to our newsletter. 

Provide your email for exclusive promotions, updates and more!

Thanks for contacting us. We'll get back to you as soon as possible.

By subscribing to our newsletter you agree to our privacy policy and will get commercial communication

Copyright © 2004-2024 General Medical Devices, Inc.

All Rights Reserved

Organizations Close to Our Hearts